Template Laboratory Battery Organizer 2017‑03‑24 16:28:35

Id2.16.756.5.30.1.1.10.4.19
ref
ch-palm-
Effective Date2017‑03‑24 16:28:35
Statusdraft DraftVersion Label2017
Namechpalm_entry_LaboratoryBatteryOrganizerDisplay NameLaboratory Battery Organizer
Description
According to IHE XD-LAB, the Laboratory Battery Organizer allows the grouping of results.
Laboratory reports MUST contain at least one result group with at least one result.
The grouping of multiple results (Observations) into multiple result groups (Laboratory Battery Organizers) is permitted.
ClassificationCDA Entry Level Template
Open/ClosedOpen (other than defined elements are allowed)
RelationshipSpecialization: template 2.16.840.1.113883.10.12.305 CDA Organizer (2005‑09‑07)
ref
ad1bbr-

Specialization: template 1.3.6.1.4.1.19376.1.3.1.4 (2016‑07‑05)
ref
ch-palm-
ItemDTCardConfDescriptionLabel
hl7:organizer
chpadotsizer
@classCode
cs1 … 1FBATTERY
@moodCode
cs1 … 1FEVN
hl7:templateId
II1 … 1Mchpadotsizer
@root
uid1 … 1F1.3.6.1.4.1.19376.1.3.1.4
hl7:id
II0 … 1An ID for this item CAN be filled for traceability.chpadotsizer
@extension
st1 … 1RMUST contain the ID itself. The ID MUST be unique within the system that issued the ID.
@root
uid1 … 1RMUST contain the OID of the system that issued the ID. OIDs of code systems, which are published in the public OID registry for the Swiss health care system (oid.refdata.ch) are REQUIRED. Others are NOT ALLOWED.
hl7:code
CD0 … 1 IF the results in this result group are associated to a codable group (e.g., LOINC Code 55429-5 for 'short blood count panel'), then the code SHOULD be specified. The code system is not further specified. LOINC (2.16.840.1.113883.6.1) or SNOMED-CT (2.16.840.1.113883.6.96) codes or values from the HL7 vocabulary ObservationType (2.16.840.1.113883.1.11.16226) SHOULD be used.
IF this result group contains any other results, then the code SHOULD be omitted.
chpadotsizer
@displayName
st1 … 1RThe displayName of the code MUST be specified for tracking purposes. However, processing is NOT ALLOWED. Instead, a processing system is to display a user readable language translation of the code from the value set.
@codeSystemName
st1 … 1RThe name of the code system MUST be specified for tracking purposes. However, processing is NOT ALLOWED.
@codeSystem
oid1 … 1RThe OID of the code system MUST be specified.
@code
cs1 … 1RThe code MUST be specified.
Included1 … 1 from 2.16.756.5.30.1.1.10.9.14 Narrative Text Reference (DYNAMIC)
The human-readable text MUST be generated automatically from the structured information of this element.
The text element MUST contain the reference to the corresponding text in the human readable part, ONLY.
hl7:text
ED1 … 1MCDA‑CH V2
hl7:reference
TEL1 … 1MThe reference to the corresponding text in the human readable part must be specified by reference to content[@ID]: reference[@value='#xxx']CDA‑CH V2
@value
1 … 1RReference to the narrative part of the section in the format '#xxx', where xxx is the ID of the corresponding element.
 Schematron assertrolered error 
 teststarts-with(@value,'#') 
 MessageThe @value attribute content MUST conform to the format '#xxx', where xxx is the ID of the corresponding 'content' element. 
 Variable letNameidvalue 
 Valuesubstring-after(@value,'#') 
 Schematron assertrolered error 
 testancestor::hl7:structuredBody//*[@ID=$idvalue] 
 MessageNo narrative text found for this reference (no content element within this document has an ID that corresponds to '<value-of select="$idvalue"/>'). 
hl7:statusCode
CS1 … 1M The status 'completed' means that all expected results for this results group are present and in a final state.
The status 'aborted' means that the examinations had to be aborted. Some results MAY be available.
chpadotsizer
 CONF
The value of @code shall be drawn from value set 1.3.6.1.4.1.19376.1.3.11.2 ActStatusAbortedCompleted (DYNAMIC)
hl7:effectiveTime
IVL_TS.CH.TZ1 … 1RPeriod of findings (physiologically relevant period) for all results in this result group.chpadotsizer
hl7:low
TS.CH.TZ1 … 1RStart of the period. Accuracy of the value: Date and time with hour and minute. If it is not known, nullFavor='UNK' must be used.chpadotsizer
 Schematron assertrolered error 
 teststring-length(@value)>11 
 MessageA precision of date and time of day with at least hour and minute is required. 
hl7:high
TS.CH.TZ1 … 1REnd of the period. Accuracy of the value: Date and time with hour and minute. If it is not known, nullFavor='UNK' must be used.chpadotsizer
 Schematron assertrolered error 
 teststring-length(@value)>11 
 MessageA precision of date and time of day with at least hour and minute is required. 
Choice0 … 1
IF this result group is based on a specimen of a non-human material, the subject element MUST be declared.
Elements to choose from:
hl7:subject
0 … 1Contains 2.16.756.5.30.1.1.10.4.5 Non-Human Subject - subject Containment (DYNAMIC)chpadotsizer
hl7:subject
0 … 1Contains 2.16.756.5.30.1.1.10.4.6 Human Patient with Non-Human Subject - subject Containment (DYNAMIC)chpadotsizer
hl7:performer
0 … *This CAN be used to indicate who has performed the test. This information supersedes any information recorded at higher level.
Contains 2.16.756.5.30.1.1.10.4.7 Laboratory Performer - performer Containment (DYNAMIC)
chpadotsizer
hl7:author
0 … *This CAN be used to indicate who has documented the result. This information supersedes any information recorded at higher level.
Contains 2.16.756.5.30.1.1.10.9.16 Author Compilation with name, addr and telecom (DYNAMIC)
chpadotsizer
Choice0 … *
This CAN be used to identify other parties:
  • Who has verified the result (AUTHEN).
  • Which external laboratory has delivered the result (RESP).
  • Which device (analyzer) was used to determine the result (DEV).
Elements to choose from:
hl7:participant
0 … *Who has verified the result (AUTHEN). This information supersedes any information recorded at higher level.
Contains 2.16.756.5.30.1.1.10.4.8 Laboratory Results Validator - participant Containment (DYNAMIC)
chpadotsizer
hl7:participant
0 … *Which external laboratory has delivered the result (RESP). This information supersedes any information recorded at higher level.
Contains 2.16.756.5.30.1.1.10.4.9 Responsible Party - participant Containment (DYNAMIC)
chpadotsizer
hl7:participant
0 … *Which device (analyzer) was used to determine the result (DEV). This information supersedes any information recorded at higher level.
Contains 2.16.756.5.30.1.1.10.4.10 Device - participant Containment (DYNAMIC)
chpadotsizer
Choice0 … *Elements to choose from:
hl7:entryRelationship
0 … *Contains 2.16.756.5.30.1.1.10.4.11 Specimen Collection (DYNAMIC)chpadotsizer
@typeCode
cs1 … 1FCOMP
hl7:entryRelationship
0 … *Contains 2.16.756.5.30.1.1.10.4.3 Laboratory Observation (DYNAMIC)chpadotsizer
@typeCode
cs1 … 1FCOMP
hl7:entryRelationship
0 … *Contains 2.16.756.5.30.1.1.10.4.18 Multimedia Embedded Content (DYNAMIC)chpadotsizer
@typeCode
cs1 … 1FCOMP
hl7:entryRelationship
0 … *Contains 2.16.756.5.30.1.1.10.4.2 Annotation Comments (DYNAMIC)chpadotsizer
@typeCode
cs1 … 1FCOMP
hl7:reference
0 … *This CAN be used to make reference to an external document.
Contains 2.16.756.5.30.1.1.10.4.29 External document (DYNAMIC)
chpadotsizer
@typeCode
cs1 … 1FREFR