Template

Show index

Template Specimen Type 2017‑03‑24 16:36:19

Id2.16.756.5.30.1.1.10.4.27
ref
ch-palm-
Effective Date2017‑03‑24 16:36:19
Statusdraft DraftVersion Label2017
Namechpalm_entry_SpecimenTypeDisplay NameSpecimen Type
DescriptionSpecimen Type.
ClassificationCDA Entry Level Template
Open/ClosedOpen (other than defined elements are allowed)
RelationshipSpecialization: template 2.16.840.1.113883.10.12.313 CDA PlayingEntity (2005‑09‑07)
ref
ad1bbr-

Specialization: template 1.3.6.1.4.1.19376.1.3.1.9999.10.9.6 (2016‑07‑05)
ref
ch-palm-
ItemDTCardConfDescriptionLabel
Choice0 … *Elements to choose from:
  • hl7:code
  • hl7:code
hl7:code
CE0 … 1RIHE XD-LAB requires coding of the specimen. However, since the laboratory results have to be coded with LOINC, the specimen is already defined via the 'System' axis of LOINC. Therefore the following, fixed code CAN be used for laboratory reports in Switzerland.chpadotsType
@codeSystemName
st1 … 1FCDA-CH Material
@codeSystem
oid1 … 1F2.16.756.5.30.2.1.1.10
@code
cs1 … 1FLOINC
hl7:code
CD0 … 1RIF the coding of the specimen with LOINC's 'System' axis is insufficient, a precise code CAN be used. The use of SNOMED-CT is recommended.chpadotsType
@displayName
st1 … 1RThe displayName of the code MUST be specified for tracking purposes. However, processing is NOT ALLOWED. Instead, a processing system is to display a user readable language translation of the code from the value set.
@codeSystemName
st1 … 1RThe name of the code system MUST be specified for tracking purposes. However, processing is NOT ALLOWED.
@codeSystem
oid1 … 1RThe OID of the code system MUST be specified.
@code
cs1 … 1RThe code MUST be specified.
hl7:qualifier
CR0 … *A precision of the code CAN be specified using a qualifier. Not further specified.chpadotsType
hl7:name
CV0 … *chpadotsType
hl7:value
CD0 … *chpadotsType